Thermometer
MODEL TES-1300/1302/1303
  • Max Hold Function
  • Data Hold Function
  • Aux output JACKS
  • T1-T2 differential measurement

  • Specifications for TES-1300/1302/1303

    Measuring Range: -50oC -1300oC;
    -50oF -1999oF
    Sensor: K Type thermocouple
    Input: Single(1300), Dual (1302,1303)
    Resolution: -50oC + 199.9oC...0.1oC
    -50oF + 199.9oF...0.1oF
    (Others-1oC/1oF)

    Accuracy:
    -50oC to 199.9oC...0.3% rdg + 1oC
    -50oF to 199.9oF...0.3% rdg + 2oF
    -50oC to 1000oC...0.5% rdg + 1oC
    -50oF to 1999oF. ..0.5% rdg + 2oF
    1000oC to 1300oC..0.75% rdg + 1oC
    (apply to mainframe only at 23oC?/FONT>5oC)
    Function : H Data Hold
    & Display oC/oF selection
    0.1o/1oresolution selection
    T1 & T2 dual points display
    T1 - T2 differential measurement
    3-1/2 digit LCD
    Overange: 1------
    Low Battery: BT

    Temperature Coefficient
    Less than 0.1 time the applicable
    accuracy specification per oC (oF)

    Input Protection
    60V DC or 24 Vrms AC maximum
    input voltage on any combination of
    input Connector.
    Sampling Rate: 2.5 times/sec.
    Operating Condition
    0oC - 50oC (32oF-122oF),
    0-90% R.H. (0-35oC)
    Power Source: one 9V battery

    Size Weight:
    135mm(L)x72mm(W)x31mm(H) & 57g
    Accessories: Battery & Instruction
    Manual
    * OPTIONAL TEMPERATURE PROBES
    ______________________________________
    Model Range(Type K(CA)THERMOCOUPLE)
    TP-K01 Bead -50oC to 200oC/-58oF to 392oF
    General Purpose
    TP-K02 Immersion -50oC to 1000oC/-58oF to 1832oF
    Liquid or gels.
    TP-K03 Surface -50oC to 750oC/-58oF to 1382oF

    TES Electrical Electronic Corp.
    No. 57 Jen-Ai Road Sec. 2 TAIPEI, TAIWAN.
    TEL : 886-2-2393-9142
    FAX : 886-2-2395-9146
    E-mail : [email protected]

    [PREVIOUS PAGE]
    [NEXT PAGE]
    [INDEX]
    [HOME]